LOGOAnalytical Chemistry
WWW.ARC.AB.CA download chinese software
leftup.gif (1559 ×Ö½Ú)

left.gif (633 ×Ö½Ú)

Applications of ICP-MS:
Air Particulate Matter
Analysis of Waters
Arsenic Speciation
Plant Tissue
Animal Tissue
Soil and Sediments
Crude Oil & Lubricants
Biological Samples
Geological Samples
greenball.gif (258 bytes)Analytical chemistry
greenball.gif (258 bytes)Inorganic Analysis
greenball.gif (258 bytes)Organic Analysis
greenball.gif (258 bytes)Pretroleum Products
arc_home.gif (258 bytes)

Please Contact:

Xinbang Feng, Ph.D.
Phone: 780 632-8464
E-mail: [email protected]

Angela Wharmby, B.Sc.
Phone: 780 632-8448
E-mail: [email protected]

Adolph Wittmeier
Phone: 780 632-8466
E-mail: [email protected]

Alan Hyrynik
Phone: 780 632-8440
E-mail: [email protected]

Our Facsimile:
780 632-8620

Airborne Particulate Matter

Airborne particulate matter samples include: fine, coarse and total suspended particulate collected from ambient and source emissions.

e1.jpg (16665 bytes)

Applications:

greenball.gif (258 bytes)Ambient air particulate matter
greenball.gif (258 bytes)Personal Exposure Assessment
greenball.gif (258 bytes)Source Apportionment
greenball.gif (258 bytes)Stack Emission

Air PM Sasmpler:

greenball.gif (258 bytes)The PEMTM Sampler
greenball.gif (258 bytes)The MINIVOL Portable Sampler
greenball.gif (258 bytes)The Dichotomous Sampler
greenball.gif (258 bytes)The Versatile Air Pollutant Sampler (VAPS)


Measurement of Mass Concentration

e2.jpg (14041 bytes)

The net PM mass collected on air filters is determined by weighing the filter before and after sampling with the micro-balance shown above. It has a minimum readability of 0.001 mg.

Gravimetric Weighing Services:
greenball.gif (258 bytes)TSP/PM10/PM2.5
greenball.gif (258 bytes)Provision of Glass Fibre, Quartz, and Teflon filters
greenball.gif (258 bytes)37 mm, 47 mm and 8X10 inch filters
greenball.gif (258 bytes)Initial and Final Weights
greenball.gif (258 bytes)Temperature and Humidity controlled Weighing Environment
greenball.gif (258 bytes)Filters Equilibrated for at least 24 Hours in the Weighing Environment prior to Weighing
greenball.gif (258 bytes)After Daily Calibration, Control Weight and Control Filter are weighed at Regular Intervals to facilitate Drift Correction
greenball.gif (258 bytes)Maintenance and Calibration with NIST traceable weights done annually by an Independent Agency
greenball.gif (258 bytes)Loading and Unloading of Filter Packs in a Location dedicated to Filter Processing
greenball.gif (258 bytes)Careful Packaging of Filters and Filter Packs for Shipment
greenball.gif (258 bytes)Additional Analyses of Exposed Filters for Inorganic and Organic Parameters readily available at the Alberta Research Council's Analytical Chemistry Business Unit


Multi-elemental Analysis by ICP-MS

(Click an element for more information)

HydrogenHeliumLithiumBerylliumBoronCarbonNitrogenOxygenFluorineNeonSodiumMagnesiumAluminumSiliconPhosphorusSulfurChlorineArgonPotassiumCalciumScandiumTitaniumVanadiumChromiumManganeseIronCobaltNickelCopperZincGalliumGermaniumArsenicSeleniumBromineKryptonRubidiumStrontiumYttriumZirconiumNiobiumMolybdenumTechnetiumRutheniumRhodiumPalladiumSilverCadmiumIndiumTinAntimonyTelluriumIodineXenonCesiumBariumLanthanumCeriumPraseodymiumNeodymiumPromethiumSamariumEuropiumGadoliniumTerbiumDysprosiumHolmiumErbiumThuliumYtterbiumLutetiumHafniumTantulumTungstenRheniumOsmiumIrdiumPlatinumGoldMercuryThaliumLeadBismuthPoloniumAstitineRadonFranciumRadiumActiniumThoriumProtactiniumUraniumNeptuniumPlutoniumAmericiumCuriumBerkeliumCaliforniumEinsteiniumFermiumMendeleviumNobeliumLawrenciumRutherfordiumHahniumSeaborgiumNeilsboriumHassiumMeitneriumElement 110Element 111Element 112Element 113

Up to 65 Elements/Metals determined for Air Filters
Analysis of Quality Control Samples including:
greenball.gif (258 bytes)Reagent Blanks
greenball.gif (258 bytes)Lab Filter Blanks
greenball.gif (258 bytes)Field Blanks
greenball.gif (258 bytes) NIST Standard Reference Materials such as SRM 1633b (Coal Fly Ash) and SRM 1648 (Urban Particulate Matter)


%Recoveries of Elements in NIST Standard Reference Materials SRM 1633b (Coal Fly Ash) and 1648 (Urban Particulate Matter)
Element M/Z Unit Certified Value 1633b %Recoveries Certified Value 1648 %Recoveries
Ag 107 µg/g - - 6 ** 98.7 ± 3, 4
Al 27 mg/g 150.5 ± 2.7 98.5 ± 1, 4 34.2 ± 1.1 103 ± 1, 4
As 75 �g/g 136.2 ± 2.6 98.0 ± 1, 4 115 ± 10 99.9 ± 3, 4
B 10 µg/g - - 158 * 103 ± 3, 4
Ba 137 µg/g 709 ± 27 99.8 ± 2, 4 737 ** 104 ± 1, 4
Be 9 µg/g - - 2.6 ± 0.4* 90.4 ± 7, 4
Ca 43 mg/g 15.1 ± 0.6 102 ± 3, 4 58.3 ± 0.6 * 102 ± 2, 4
Cd 114 µg/g 0.784 ± 0.006 95.0 ± 4, 4 75 ± 7 97.2 ± 1, 4
Co 59 µg/g 50 ** 98.2 ± 2, 4 18 ** 102 ± 3, 4
Cr 53 µg/g 198.2 ± 4.7 102 ± 3, 4 403 ± 12 31.6 ± 1, 4
Cu 65 µg/g 112.8 ± 2.6 101 ± 2, 4 609 ± 27 97.8 ± 1, 4
Fe 57 mg/g 77.8 ± 2.3 97.3 ± 2, 4 39.1 ± 1.0 99.6 ± 2, 4
K 39 mg/g 19.5 ± 0.3 79.5 ± 7, 4 10.5 ± 0.1 91.0 ± 3, 4
Mg 25 mg/g 4.82 ± 0.08 98.6 ± 2, 4 8 ** 98.2 ± 6, 4
Mn 55 µg/g 131.8 ± 1.7 109 ± 1, 4 860 ** 100 ± 1, 4
Mo 98 µg/g - - 18.2 ± 1.9* 102 ± 2, 4
Na 23 mg/g 2.01 ± 0.03 98.4 ± 2, 4 4.25 ± 0.02 98.2 ± 1, 4
Ni 60 µg/g 120.6 ± 1.8 102 ± 4, 4 82 ± 3 102 ± 4, 4
Pb 208 µg/g 68.2 ± 1.1 101 ± 2, 4 6550 ± 80 103 ± 1, 4
Sb 121 µg/g 6 ** 90 ± 3, 4 45 ** 96.8 ± 1, 4
Se 77 µg/g 10.26 ± 0.17 102 ± 4, 4 27 ± 1 98.5 ± 13, 4
Sr 88 µg/g 1041 ± 14 101 ± 2, 4 207 ± 15 * 101 ± 1, 4
Th 232 µg/g 25.7 ± 1.3 98.7 ± 2, 4 7.4 ** 97.6 ± 2, 4
Tl 205 µg/g 5.9 ** 97.8 ± 9, 4 1.27 ± 0.06 98.1 ± 2, 4
U 238 µg/g 8.79 ± 0.36 104 ± 2, 4 5.5 ± 0.1 102 ± 2, 4
V 51 µg/g 295.7 ± 3.6 103 ± 1, 4 140 ± 3 101 ± 1, 4
Zn 66 µg/g 210 ** 112 ± 3, 4 4760 ± 140 106 ± 2, 4
** Non-certified Values
* Consensus Values
Reference:
  1. Ambient Air Particulate Matter Sampling and Multi-Elemental Measurement by Inductively Coupled Plasma-Mass Spectrometry
    Xinbang Feng, Angela Wharmby and Adolph Wittmeier,
    Recent Advances in the Science and Management of Air Toxics, Banff Rimrock Hotel, Banff, Alberta, Canada, April 9-12, 2000.

  2. Multi-Elemental Determinations of Airborne Particulate Matter by Inductively Coupled Plasma-Mass Spectrometry
    Xinbang Feng, Angela Wharmby and Adolph Wittmeier,
    46th International Conference on Analytical Sciences and Spectroscopy (46th ICASS), University of Winnipeg, Winnipeg, Manitoba, Canada, August 13-16, 2000.

  3. Gravimetric Measurement of Fine Particulate Matter and Multi-Elemental Analysis by Inductively Coupled Plasma-Mass Spectrometry
    Xinbang Feng, Angela Wharmby, Adolph Wittmeier and Shaole Wu,
    10th Annual Conference of the International Society of Exposure Analysis (ISEA2000) in Asilomar Conference Center, Monterey Peninsula, California, USA, October 24-27, 2000.



Back to Heavy Metal Analysis

Copyright © Alberta Research Council, All rights reserved.

Address:
Alberta Research Council
Analytical Chemistry
Highway 16A, 75 Street
P.O. Bag 4000, Vegreville, Alberta
Canada T9C 1T4
Tel: 780 632-8464
Fax: 780 632-8620
We welcome feedback and comments at E-mail: [email protected]
Best viewed with 800*600 pixels,16 Bit Color